β-Hydroxybutyrate dehydrogenase decorated MXene nanosheets for the amperometric determination of β-hydroxybutyrate
- Koyappayil, Aneesh;
- Chavan, Sachin Ganpat;
- Mohammadniaei, Mohsen;
- Go, Anna;
- Hwang, Sei Young;
- ... Lee, Min-Ho
WEB OF SCIENCE
54SCOPUS
64초록
MXene nanosheets of type Ti3C2Tx were modified with beta-hydroxybutyrate dehydrogenase and then used as a biosensor for amperometric sensing of beta-hydroxybutyrate. The MXene and the nanocomposite were characterized by X-ray photoelectron spectroscopy, field-emission scanning electron microscopy, X-ray diffraction, and Fourier transform infrared spectroscopy. The MXene has a layered structure and proved to be an excellent immobilization matrix providing good compatibility with the enzyme beta-hydroxybutyrate dehydrogenase. The MXene-based biosensor, best operated at a potential of - 0.35 V (vs. Ag/AgCl), displays a wide linear range (0.36 to 17.9 mM), a sensitivity of 0.480 mu A mM(-1) cm(-2), and a low detection limit (45 mu M). The biosensor was successfully applied to the determination of beta-hydroxybutyrate in (spiked) real serum samples.