Quantitative Raman analysis of microplastics in water using peak area ratios for concentration determination

  • Jung, Eun Su; 
  • Choe, Jin Hyun; 
  • Kim, Jin Seok; 
  • Ahn, Da Won; 
  • Yoo, Jinuk; 
  • ... Pyo, Sung Gyu; 
  • 외 1명
Citations

WEB OF SCIENCE

36
Citations

SCOPUS

38

초록

This study presents a novel analytical method for the quantitative and qualitative analysis of microplastics (MPs) in deionized (DI) water using Raman spectroscopy. Raman peak area ratios of 1295 cm-1 for polyethylene (PE) and 637 cm-1 for polyvinylchloride (PVC) to the broad H2O peak were utilized to establish a calibration model for the concentration of MPs dispersed in DI water at 0.1 wt% to 1.0 wt%. The calibration model demonstrated R2 values of 0.98537 for PE and 0.99511 for PVC, indicating high linearity between the peak area ratio and concentration. The calibration model was validated using mixed PE and PVC samples to confirm its applicability to real-world water bodies, where multiple types of MPs are present. The calculated standard error of calibration (SEC) and relative standard error of calibration (%RSEC) values further confirmed the accuracy of the predictions, providing a robust approach for detecting and quantifying MPs in aquatic environments.

키워드

ENVIRONMENTAL-SAMPLES; SPECTROSCOPY; SEA
제목
Quantitative Raman analysis of microplastics in water using peak area ratios for concentration determination
저자
Jung, Eun Su; Choe, Jin Hyun; Kim, Jin Seok; Ahn, Da Won; Yoo, Jinuk; Choi, Tae Min; Pyo, Sung Gyu
DOI
10.1038/s41545-024-00397-4
발행일
2024-10
유형
Article
저널명
NPJ CLEAN WATER
권
7
호
1

파일 다운로드

Thumbnail