Performance Evaluation and Measurement Uncertainty Determination of the New Version of the Abbott ARCHITECT 25-OH Vitamin D 5P02 Assay
- Lim, Yong Kwan;
- Park, Ae Ja;
- Kweon, Oh Joo;
- Choi, Jee-Hye
WEB OF SCIENCE
6SCOPUS
5초록
Objectives We evaluated performance and measurement uncertainty of the newly released ARCHITECT 25-OH vitamin D 5P02 assay (Abbott Diagnostics, Abbott Park, IL). Methods In total, 300 samples were used to compare the results from ELECSYS Vitamin D Total (Roche Diagnostics, Mannheim, Germany) and ADVIA Centaur Vitamin D Total (Siemens, Tarrytown, NY). To quantify the measurement uncertainty, 25 samples and four levels of standard reference material (SRM) were measured using isotope dilution-liquid chromatography-tandem mass spectrometry according to international guidelines. Results The results of the ARCHITECT assay were equivalent to other immunoassays, but correlation coefficients were lower than the recommended criterion. SRM level 2 was considered adequate for uncertainty estimation, and the expanded measurement uncertainty of the ARCHITECT assay was 4.2%, which was superior to the other two assays. Conclusions The restandardized ARCHITECT assay has acceptable performance in a clinical setting. However, there is still a need for further standardization of total vitamin D measurement among the automated immunoassays.
키워드
- 제목
- Performance Evaluation and Measurement Uncertainty Determination of the New Version of the Abbott ARCHITECT 25-OH Vitamin D 5P02 Assay
- 저자
- Lim, Yong Kwan; Park, Ae Ja; Kweon, Oh Joo; Choi, Jee-Hye
- 발행일
- 2019-02
- 유형
- Article
- 권
- 151
- 호
- 2
- 페이지
- 209 ~ 216
- 언어
- ENG
- 출판사
- OXFORD UNIV PRESS INC
- 발행국가
- 미국
- 분량
- 8 페이지
- ISSN
- E 1943-7722
P 0002-9173