Reconfigurable RF receiver using compact distributed switch matrix for array coils

Citations

WEB OF SCIENCE

0
Citations

SCOPUS

0

초록

Magnetic resonance (MR) signals captured by RF array coils should be delivered to analogue-to-digital converter (ADC) channels through an independent receiver (Rx) circuitry for precise MRI process. Since different types of RF coils should be utilised by a fixed port configuration on a patient table with a limited number of ADC channels, a highly flexible Rx front-end design is necessary. In this Letter, a compact distributed switch matrix structure for multi-channel RF coil Rx is proposed for supporting different types of array coils with small circuit size and sufficient reconfigurability. The proposed structure is configured by a multiple number of unit matrix cells consisting of a single double pole double throw and two single pole double throw RF switches to control the pathways of received MR signals. The proposed structure was implemented and verified at 9.4-Tesla (9.4 T) MR frequency (400 MHz) by comparing different RF path losses.

키워드

MRI
제목
Reconfigurable RF receiver using compact distributed switch matrix for array coils
저자
Lee, H. L.
DOI
10.1049/el.2018.6579
발행일
2019-02
유형
Article
저널명
Electronics Letters
권
55
호
3
페이지
137 ~ +