Analysis of position error by time constant in read-out resistive network for gamma-ray imaging detection system

  • Jeon, Su-Jin; 
  • Park, Chang-In; 
  • Son, Byung-Hee; 
  • Jung, Mi; 
  • Jang, Teak-Jin; 
  • ... Choi, Young-Wan; 
  • 외 1명
Citations

WEB OF SCIENCE

1
Citations

SCOPUS

1

초록

Position-sensitive photomultiplier tubes (PSPMTs) in array are used as gamma ray position detector. Each PMT converts the light of wide spectrum range (100 nm similar to 2500 nm) to electrical signal with amplification. Because detection system size is determined by the number of output channels in the PSPMTs, resistive network has been used for reducing the number of output channels. The photo-generated current is distributed to the four output current pulses according to a ratio by resistance values of resistive network. The detected positions are estimated by the peak value of the distributed current pulses. However, due to parasitic capacitance of PSPMTs in parallel with resistor in the resistive network, the time constants should be considered. When the duration of current pulse is not long enough, peak value of distributed pulses is reduced and detected position error is increased. In this paper, we analyzed the detected position error in the resistive network and variation of time constant according to the input position of the PSPMTs.

키워드

Gamma-ray imaging detection system; resistive network; detected position error; time constant; ARRAYS
제목
Analysis of position error by time constant in read-out resistive network for gamma-ray imaging detection system
저자
Jeon, Su-Jin; Park, Chang-In; Son, Byung-Hee; Jung, Mi; Jang, Teak-Jin; Lee, Chun-Sik; Choi, Young-Wan
DOI
10.1117/12.2212515
발행일
2016-03
유형
Proceedings Paper
저널명
SMART PHOTONIC AND OPTOELECTRONIC INTEGRATED CIRCUITS XVIII
권
9751