Efficacy and tolerability of metallic stent in patients with malignant prostatic obstruction secondary to prostatic cancer

  • Choi, Se Young
  • Lim, Bumjin
  • Chi, Byung Hoon
  • Kim, Jung Hoon
  • Lee, Wonchul
  • 외 4명
Citations

WEB OF SCIENCE

3
Citations

SCOPUS

3

초록

Objective To evaluate the efficacy and tolerability of a self-expandable covered metallic stent in patients with malignant prostatic obstruction secondary to prostate cancer (PC). Methods We reviewed 22 cases of insertion of self-expandable covered metallic stents with barbs. Data were collected about PC status. Uroflowmetry variables, residual urine volume, International Prostate Symptom Score (IPSS), quality of life (QOL), and duration from stent insertion to removal were surveyed. These clinical parameters were compared before and after stent insertion. Results The patients with PC showed a mean age of 75.5 +/- 6.5 years and mean 5.1 +/- 1.9 Charlson comorbidity index. The average flow rate (2.4 +/- 1.9 vs 5.9 +/- 2.4 mL/s, P = .005), peak flow rate (6.9 +/- 6.2 vs 14.1 +/- 5.5 mL/s, P = .003), flow time (54.6 +/- 29.1 vs 23.6 +/- 13.7 s, P = .002), residual urine volume (178.7 +/- 195.5 vs 7.0 +/- 7.1 mL, P = .004), IPSS (26.2 +/- 8.1 vs 8.0 +/- 6.5 points, P = .001), and QOL (4.7 +/- 1.3 vs 2.4 +/- 2.1 points, P = .030) improved between before and after stent insertion, respectively. Pain was the most common complication, but 60% of the patients were managed without any intervention. There were hematuria, urinary retention, urinary frequency, obstruction, and urinary incontinence. However, there was no urinary tract infection due to the stent. The median time to stent removal was 5.7 months. Conclusions The stent was maintained for about 6 months with improved objective and subjective outcomes. The patients with PC, who had a poor comorbidity index and advanced PC status showed a tolerable maintenance period. Self-expandable covered metallic stents can be used for PC patients with a short life expectancy and unsuitability for general anesthesia.

키워드

barbprostate cancerself‐expandable covered metallic stentstransurethral resection of the prostateurethralobstruction
제목
Efficacy and tolerability of metallic stent in patients with malignant prostatic obstruction secondary to prostatic cancer
저자
Choi, Se YoungLim, BumjinChi, Byung HoonKim, Jung HoonLee, WonchulKyung, Yoon SooYou, DalsanSong, Ho-YoungKim, Choung-Soo
DOI
10.1111/luts.12367
발행일
2021-07
유형
Article
저널명
LUTS: Lower Urinary Tract Symptoms
13
3
페이지
329 ~ 334