Accelerating acquisition of readout-segmented echo planar imaging (rs-EPI) with a simultaneous multislice (SMS) technique for diffusion-weighted (DW) breast MRI: Evaluation of image quality factors
Citations

WEB OF SCIENCE

4
Citations

SCOPUS

3

초록

Purpose: This study aims to compare the image quality, apparent diffusion coefficient (ADC), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) values, and scan time between readout-segmented echo planar imaging (rs-EPI) and simultaneous multislice (SMS) rs-EPI sequences. Methods: A total of 80 consecutive women who underwent breast diffusion-weighted imaging (DWI) were included, and two rs-EPI DWI sequences with and without SMS were acquired and compared. Qualitative analysis involved three radiologists independently scoring image quality and radiologist preference. For quantitative comparison, the radiologists independently measured the ADC values in patients, while SNR, CNR, and ADC values were measured on a phantom. Results: The acquisition time was 5:47 min for rs-EPI and 3:20 min for SMS rs-EPI. In terms of image quality, scores were similar between rs-EPI and SMS rs-EPI sequences in the pooled data set, with the exception of skin-line distinction (p = 0.001) and background noise (p < 0.001). All radiologists considered SMS rs-EPI as equal or superior to rs-EPI in more than 70 % of cases. SMS rs-EPI demonstrated significantly higher ADC values than rs-EPI by all radiologists (p ≤ 0.002). For the phantom measurement, ADC (SMS: 1.26 ± 0.68 and RS: 1.26 ± 0.68, p = 0.198), SNR (SMS: 540.6 ± 342.1 and RS: 558.8 ± 523.2, p = 0.927), and CNR (SMS: 235.5 ± 38.9 and RS: 252.8 ± 108.0, p = 0.784) values did not significantly differ between the two sequences. Conclusion: SMS rs-EPI exhibited comparable image quality and similar ADC, SNR, and CNR values to rs-EPI while reducing the scan time by 42%. © 2023 Elsevier B.V.

키워드

Breast neoplasmsDiffusion magnetic resonance imagingImage qualityMagnetic resonance imaging
제목
Accelerating acquisition of readout-segmented echo planar imaging (rs-EPI) with a simultaneous multislice (SMS) technique for diffusion-weighted (DW) breast MRI: Evaluation of image quality factors
저자
Ahn, Hye ShinKim, Sung HunKim, Ji YounHong, Min JiLee, Hyun-Soo
DOI
10.1016/j.ejrad.2023.111251
발행일
2024-01
유형
Article
저널명
European Journal of Radiology