Enhanced detection of single-cell-secreted proteins using a fluorescent immunoassay on the protein-G-terminated glass substrate

Citations

WEB OF SCIENCE

7
Citations

SCOPUS

6

초록

We present an evaluation of protein-G-terminated glass slides that may contain a suitable substrate for aligning the orientation of antibodies to obtain better binding moiety to the target antigen. The results of the protein-G-terminated slides were compared with those obtained with epoxy-based slides to evaluate signal enhancement for human immunoglobulin G (IgG) targets, and an increase in the average fluorescence intensity was observed for the lowest measurable amount of IgG target in the assay using protein-G-terminated slides. Applying this strategy for signal amplification to single-cell assays improves the limits of detection for human IgG protein and cytokines (interleukin-2 and interferon-gamma) captured from hybridomas. Our data indicate that protein-G-terminated slides have a higher binding capacity for antigens and have better spot-to-spot consistency than that of traditional epoxy-based slides. These properties would be beneficial in the detection of fine amounts of single-cell-secreted proteins, which may provide key insights into cell-cell communication and immune responses.

키워드

microwell array; antibody's orientation; single cell analysis; secreted cytokine; protein-G-terminated surface; ANTIBODY IMMOBILIZATION TECHNIQUES; FC FRAGMENT; MICROARRAYS; BINDING; SURFACE; DNA; COMPLEX; TOOL
제목
Enhanced detection of single-cell-secreted proteins using a fluorescent immunoassay on the protein-G-terminated glass substrate
저자
Jeong, Yoon; Lee, Kwan Hong; Park, Hansoo; Choi, Jonghoon
DOI
10.2147/IJN.S92596
발행일
2015-11
유형
Article
저널명
International Journal of Nanomedicine
권
10
호
1
페이지
7197 ~ 7205

파일 다운로드