Validation of the COSINE-100U NaI(Tl) encapsulation for low-temperature operation in liquid scintillator
  • Park, K.
  • Cho, S.J.
  • França, L.E.
  • Ha, C.
  • Kim, J.
  • 외 10명
Citations

SCOPUS

0

초록

The COSINE-100U (upgrade) will enhance the sensitivity of the COSINE-100 dark matter search by operating the detector array immersed in liquid scintillator (LS) at -30°C. To validate the detector design for these conditions, we constructed a module using the COSINE-100U encapsulation and performed a dedicated long-term stability study. The module was first monitored at room temperature for ∼110 days in air, followed by a one-week immersion in LAB-based LS to verify initial compatibility. Upon confirming stable optical performance, the temperature was lowered to -33°C. During approximately 150 days of continuous operation at low temperature, we observed no degradation in performance. These results demonstrate the chemical and mechanical robustness of the encapsulation, confirming its suitability for the COSINE-100U physics run.

키워드

Dark Matter detectors (WIMPs, axions, etc.)Detector cooling and thermo-stabilizationDetector design and construction technologies and materialsScintillators, scintillation and light emission processes (solid, gas and liquid scintillators)
제목
Validation of the COSINE-100U NaI(Tl) encapsulation for low-temperature operation in liquid scintillator
저자
Park, K.Cho, S.J.França, L.E.Ha, C.Kim, J.Kim, K.W.Kim, S.H.Kim, W.K.Ko, Y.J.Lee, D.H.Lee, H.S.Lee, I.S.Lee, S.H.Park, S.D.Yu, G.H.
DOI
10.1088/1748-0221/21/03/T03010
발행일
2026-03