A microfluidic device with integrated impedance detection for λ-DNA

  • Park, Myung-Il; 
  • Hong, Jongin; 
  • Yoon, Dae Sung; 
  • Park, Chong-Ook; 
  • Im, Geunbae
Citations

WEB OF SCIENCE

1

초록

The large optical detection systems that are typically utilized at present may not be able to reach their full potential as portable analysis tools. Accurate, early, and fast diagnosis for many diseases requires the direct detection of biomolecules such as DNA, proteins, and cells. In this research, a glass microchip with integrated microelectrodes has been fabricated, and the performance of electrochemical impedance detection was investigated for the biomolecales. We have used label-free lambda-DNA as a sample biomolecule. By changing the distance between microelectrodes, the significant difference between DW and the TE buffer solution is obtained from the impedance-frequency measurements. In addition, the comparison for the impedance magnitude of DW, the TE buffer, and lambda-DNA at the same distance was analyzed.

제목
A microfluidic device with integrated impedance detection for λ-DNA
저자
Park, Myung-Il; Hong, Jongin; Yoon, Dae Sung; Park, Chong-Ook; Im, Geunbae
DOI
10.1557/PROC-773-N6.5
발행일
2003
유형
Proceedings Paper
저널명
BIOMICROELECTROMECHANICAL SYSTEMS (BIOMEMS)
권
773
페이지
149 ~ 154