상세 보기
Real-Time Retinal Imaging with a Parallel OCT Using a CMOS Smart Array Detector
- jeehyun kim;
- Thomas Milner;
- Sohn, Bong Soo
초록
We present a parallel optical coherence tomography(OCT) system using a complementary metal-oxide-semiconductor(CMOS) smart array detector. This paper includes descriptions of the CMOS smart array detector, and the experimental results of scattering surface targets including a wet tissue on an Air Force target, and {\it ex vivo} gold fish retina where the boundary between the vitreous humor and the ganglion cell layer could be identified in the acquired images. After three-dimensional volume rendering, several other layers between the ganglion cell layer and the choroid layer could be observed even though the contrast between the layers was weak. Control of the current version of the CMOS smart array detector requires significant time for maintenance. Although the current CMOS smart array detector needs improvement, the design is a promising ophthalmologic instrument and can provide fast retinal scanning and prompt display of the acquired images.
키워드
- 제목
- Real-Time Retinal Imaging with a Parallel OCT Using a CMOS Smart Array Detector
- 제목 (타언어)
- Real-Time Retinal Imaging with a Parallel OCT Using a CMOS Smart Array Detector
- 저자
- jeehyun kim; Thomas Milner; Sohn, Bong Soo
- 발행일
- 2007-11
- 권
- 51
- 호
- 5
- 페이지
- 1787 ~ 1791
- 언어
- ENG
- 출판사
- 한국물리학회
- 발행국가
- 대한민국
- 분량
- 5 페이지
- ISSN
- E 1976-8524
P 0374-4884