상세 보기
Bio-mechanical analysis of a fractured tibia with composite bone plates according to the diaphyseal oblique fracture angle
- Kim, Hyun-Jun;
- Kim, Suk-Hun;
- Chang, Seung-Hwan
WEB OF SCIENCE
62SCOPUS
67초록
The flexible fixation method is known to provide beneficial conditions for healing bone fractures by allowing appropriate relative movement at the fracture site. Using finite element analysis (FEA), we investigated the healing efficiency of flexible composite bone plates applied to a tibia with diaphyseal oblique fractures. To construct the finite element model, the time-dependent properties of living tissue (callus) were estimated using healing rates that were updated at every healing period using iterative calculations of the interfragmentary strain distributions according to oblique angles and plate properties. The calculation results showed that the flexible composite bone plates (Twintex [0](18) and WSN3k [+/-30](9)) provided high healing rates compared to conventional steel plates. (C) 2011 Elsevier Ltd. All rights reserved.
키워드
- 제목
- Bio-mechanical analysis of a fractured tibia with composite bone plates according to the diaphyseal oblique fracture angle
- 저자
- Kim, Hyun-Jun; Kim, Suk-Hun; Chang, Seung-Hwan
- 발행일
- 2011-06
- 유형
- Article
- 권
- 42
- 호
- 4
- 페이지
- 666 ~ 674
- 언어
- ENG
- 출판사
- ELSEVIER SCI LTD
- 발행국가
- 영국
- 분량
- 9 페이지
- ISSN
- E 1879-1069
P 1359-8368