인조골에 체결된 유리섬유/폴리프로필렌 복합재료 고정판의수분 환경 피로 특성

Fatigue Characterization of Glass/Polypropylene Composite Bone Plates Locked with an Artificial Tibia under Moisture Environment

초록

In this study, bone plate made of glass/polypropylene composite material which was applied to an artificial bone was tested to check the service ability under fatigue loading. To check serviceability of composite bone plates fatigue test was carried out considering changes in the moisture absorption rate, locking position of screws and loading condition. Test results showed that all the tested specimens had the fatigue life more than one million cycles which was much higher fatigue life than the expected value of 650,000 cycles. Screw position was not critical impact on the deformation of the fracture site. In this paper, the mechanical performance of the glass/polypropylene composite was verified by fatigue test under various water absorption conditions, and this result may give useful information on the design of composite bone plate.

키워드

유리섬유/폴리프로필렌 복합재료; 피로실험; 고정판; 수분 흡수; glass/polypropylene composite; fatigue test; bone plate; water absorption
제목
인조골에 체결된 유리섬유/폴리프로필렌 복합재료 고정판의수분 환경 피로 특성
제목 (타언어)
Fatigue Characterization of Glass/Polypropylene Composite Bone Plates Locked with an Artificial Tibia under Moisture Environment
저자
한민구; 장승환
DOI
10.7234/composres.2013.26.5.328
발행일
2013-10
저널명
Composites Research
권
26
호
5
페이지
328 ~ 333

파일 다운로드

Thumbnail