Background: Estimating measurement uncertainty (MU) for high-sensitivity cardiac troponin I (hscTnI) is challenging due to complex calibration requirements. We aimed to quantify the MU of hscTnI (Abbott, IL, USA) using Monte Carlo simulation (MCS), a computational algorithm that simulates statistical sampling to achieve numerical results through rigorous mathematical calculations. Methods: Precision profiling of the instrumental signal across the analytical range was performed using internal quality control data. Subsequently, four-parameter logistic-log calibration models were simulated in each MCS trial. The MU of hs-cTnI corresponding to each instrumental signal was estimated using MCS. Results: As hs-cTnI concentrations increased, the MU gradually decreased, reaching its lowest value of 15.3% at 577.9 ng/L. MU values at clinical decision cutoff points were 23.6% (15.6 ng/L for women) and 20.0% (34.2 ng/L for men). Conclusions: Using MCS, we estimated the MU of hs-cTnI across its entire measurement range, with MU values ranging approximately from 15% to 25% depending on the hs-cTnI concentration. The results should be interpreted with caution, especially near clinical decision points where uncertainty may be higher.

키워드

Measurement uncertainty; Monte Carlo simulation; High sensitivity cardiac troponin I; Clinical decision points
제목
Estimating the measurement uncertainties of the Abbott Alinity i high-sensitivity cardiac troponin I assay through Monte Carlo simulation
저자
Kim, Hongkyung; Kweon, Oh Joo; Yun, Jiwon; Choe, Kye Won; Yoon, Sumi; Kim, Seung Hyeon; Lee, Mi-Kyung; Lim, Yong Kwan
DOI
10.1016/j.heliyon.2026.e45334
발행일
2026-09
유형
Article
저널명
Heliyon
권
12
호
14