In this study, we used square-type organic scintillating fibers and polymethyl methacrylate phantoms to develop an all-in-one phantom dosimeter measure the multidimensional dose distributions for a Co-60 radiotherapy source. Multidimensional dose distributions, which include the transverse, depth, and surface dose distributions, can be obtained by measuring scintillating light simultaneously in real time. An additional phantom for quality assurance before radiotherapy is not required because square-type organic scintillating fibers are directly inserted into the polymethyl methacrylate phantom. Multidimensional dose distributions can also be measured simultaneously by changing the arrangement of one-dimensional phantom dosimeters, which are very thin and can be stacked, similar to books placed on a shelf. Therefore, the proposed dosimeter can be applied as an all-in-one phantom dosimeter with a variable structure that can respond to changes in various radiotherapy treatment environments. Through experiments, the multidimensional dose distributions for the Co-60 radiotherapy unit were measured to evaluate the performance of the proposed all-in-one phantom dosimeter. The results were compared and evaluated using external beam therapy films and Monte Carlo simulations. Based on the results of this study, significant time is expected to be saved during dose measurements for quality assurance in radiotherapy, as simultaneous measurements can yield multidimensional data on the transverse, depth and surface dose distributions in real time.

키워드

All-in-one phantom dosimeter; Organic scintillating fiber; Co-60 radiotherapy unit; Multidimensional dose distribution; Quality assurance
제목
Development of all-in-one phantom dosimeter to measure multidimensional dose distribution for Co-60 radiotherapy unit
제목 (타언어)
Development of all-in-one phantom dosimeter to measure multidimensional dose distribution for 60Co radiotherapy unit
저자
Shin, Hyun Young; Kim, Jin Ho; Song, Si Won; Kim, Jinhong; Lee, Bongsoo; Moon, Joohyun; Pyeon, Cheol Ho
DOI
10.1007/s40042-021-00129-8
발행일
2021-05
유형
Article
저널명
Journal of the Korean Physical Society
권
78
호
9
페이지
829 ~ 836