상세 보기
Effect of composite bone plates on callus generation and healing of fractured tibia with different screw configurations
- Mehboob, Ali;
- Chang, Seung-Hwan
WEB OF SCIENCE
37SCOPUS
40초록
In this paper, finite element analysis of a fractured tibia with a glass/polypropylene composite implant is introduced. A rejection coefficient algorithm (for callus development) that is sensitive to interfragmentary movement is programmed, calibrated (using experimental in vivo statistics), and successfully implemented on a 3D fractured tibia model. A biphasic mechano-regulation algorithm is implemented to verify healing status under five different screw configurations (C1-C5) using glass/polypropylene composite bone plates and the development of tissue phenotypes in calluses is estimated. A 300% increase in circumferential callus volume is obtained when using the composite bone plate. Furthermore, the C5 configuration of the composite bone plate results in a maximum interfragmentary movement of 4.33% on day one with faster and stronger healing through 95% of bone growth during the final day of healing.
키워드
- 제목
- Effect of composite bone plates on callus generation and healing of fractured tibia with different screw configurations
- 저자
- Mehboob, Ali; Chang, Seung-Hwan
- 발행일
- 2018-10
- 유형
- Article
- 권
- 167
- 페이지
- 96 ~ 105
- 출판사
- ELSEVIER SCI LTD
- 발행국가
- 영국
- 분량
- 10 페이지
- ISSN
- E 1879-1050
P 0266-3538