상세 보기
The Use of a 20-Gauge Valved Cannula During Pars Plana Phacofragmentation With a 23-Gauge Ultrasonic Fragmatome
- Kim, Jee Taek;
- Eom, Youngsub;
- Ahn, Jaemoon;
- Kim, Seong-Woo;
- Huh, Kuhl
WEB OF SCIENCE
3SCOPUS
3초록
BACKGROUND AND OBJECTIVE: To evaluate the usefulness of a 20-gauge cannula to maintain a self-sealing sclerotomy wound after 23-gauge phacofragmentation. PATIENTS AND METHODS: This retrospective study compared the suture rates after 23-gauge phacofragmentation when the 23-gauge cannula was temporarily replaced with a 20-gauge valved metal cannula versus when the 23-gauge fragmatome was inserted at the sclerotomy site without a cannula. RESULTS: Whereas a sclerotomy was sutured in all 31 eyes in the without-cannula group, only one eye of 14 in the cannula group required a sclerotomy suture (P < .0001). However, there was one case of fragmatome tip fracture during fragmentation in the cannula group. CONCLUSION: A self-sealing sclerotomy wound can be maintained without suturing by using a 20-gauge metal cannula, but fragmatome tip fracture can occur during fragmentation.
키워드
- 제목
- The Use of a 20-Gauge Valved Cannula During Pars Plana Phacofragmentation With a 23-Gauge Ultrasonic Fragmatome
- 저자
- Kim, Jee Taek; Eom, Youngsub; Ahn, Jaemoon; Kim, Seong-Woo; Huh, Kuhl
- 발행일
- 2014-05
- 유형
- Article
- 권
- 45
- 호
- 3
- 페이지
- 207 ~ 210
- 언어
- ENG
- 출판사
- SLACK INC
- 발행국가
- 미국
- 분량
- 4 페이지
- ISSN
- E 2325-8179
P 2325-8160